Bowie Dick Sensor
Bowie Dick Sensor
This highly compact Bowie Dick Sensor is the only electronic test device on the market designed for 121 °C processes – and, when necessary, 134 °C. The sensor provides far more insight into critical sterilization parameters than previously possible, making it a critical tool for any steam penetration study.
Unlike traditional chemical indicators, where you subjectively decide according to a change in color, this device provides you with an accurate report that eliminates any guess work.
The device is extremely easy to use and requires little to no training. With just a single click in the software, the unit will start to measure and record data. Once the Bowie Dick Test Program is done, the device is simply placed back in the reader station where the software automatically processes the data and provides a test result – simple and easy!
The TrackSense Pro data logger series are equipped with an Interchangeable sensor option, unlocking several measuring parameters:
✓ Temperature
✓ Pressure
✓ Conductivity
✓ Relative humidity
✓ Vacuum
✓ Steam Penetration
The data loggers can easily be activated and read by the Multi reader station.
Features
– accurate, reliable and cost efficient
– very compact size
– user-friendly and intuitive software
– eliminates guesswork and provides an objective result
– replaceable PCD to run several cycles in a sequence
– verification of critical sterilization parameters beyond that of traditional methods
Sensor Specification for this configuration: Bowie Dick Sensor
ISO Standard Compliance: 11140-4:2007 type B1, B2 and B3 at 121 °C and 134 °C
3rd Party Test Institute: SAL GmbH
Type of Sterilizers: Steam Sterilizers / Autoclaves qualified for sterilization at
121°C or +134 °C according to EN 285 (volume larger than
60 L) and EN ISO 17665 (moist heat autoclaves)
Measuring Principle: Piezoresistive / Electrical Resistance
Sensor Type: Strain Gauge / Pt1000
Temperature Measuring Range: 0 to +140 °C (Calibrated +25 to +140 °C)
Temperature Accuracy: +25 to +140 °C: ± 0.05 °C
Temperature Response Time:
T-63%: 0.8 Second
T-90%: 1.7 Second
Pressure Measuring Range: 10 mBar to 6 Bar
Accuracy: ± 0.25% Full Scale (± 15 mBar)
PCD Housing Material: PEEK
Logger Specification for this configuration: Pro 3G
Operating Temperature: -20 to +150 °C
Operating Pressure: 0.001 mBar to 10 Bar ABS
Housing Material: 316L Stainless Steel
Diameter: 25 mm
Length: 100.2 mm including PCD (30 mm)
Weight with Battery: 208 g
Memory Capacity: 120,000 Data Points / 30,000 Samples
Minimum Sample Rate: 1 Second
Maximum Sample Rate: 24 Hours
Intrinsically Safe: Ex II1GD ia IIC T3 Ga, -50 °C ≤ Tamb ≤ +105 °C
Time Accuracy: ± 5 Seconds Per 24 Hours
Battery: TSP Standard Battery